High-duty lamp lights up on the PEAK N2 generator. - WKB96765

Article number: 96765

ENVIRONMENT

  • High-duty lamp lights up on the PEAK N2 generator.

ANSWER

Why high-duty lamp is on

When there is a leak somewhere and the device is not turned off and has been running for 8 hours, the lamp lights up.

 

ADDITIONAL INFORMATION

The service lamp lights up 1 year later independent of whether on or off, or after 5000 hours of running

Needs maintenance after 5000 hours

6mm tube is thinner than 1/4 inch tube

Install two 02-4600 joints (screw side is 1/4 inch, tube side is 6 mm, both outlets are thin type of 6 mm )

Nitrogen tube 00-1266 is thick 1/4 made of Teflon

Waste tube 00-1282 is thin 6 mm made of PE with dull color

 