Extremely poor sensitivity with Xevo TQD - WKB95703
Article number: 95703
SYMPTOMS
- Extremely poor sensitivity with Xevo TQD
- It happened after the instrument restart.
- In TWAVE Diagnostic, the readback of the Collision Ion Guide Pulse Amp becomes a 3-digit value and almost 0 in some places repeatedly.
- In TWAVE Diagnostic, the readback of HT supply and Collision Guide RF goes up or down repeatedly.
- Turning off the Operate doesn’t improve.
- LEDs on the TWAVE generator body are displayed in red.
ENVIRONMENT
- XevoTQD
CAUSE
Malfunction of TWAVE generator
FIX or WORKAROUND
- Improved by unplugging/plugging the LAN cable connected to the bottom
ADDITIONAL INFORMATION
In case of recurrence, replace the 700003510: 48V TWIG ANALYSER UNIT
id95703, XEVOTQD, XEVOTQDIVD