Article number: 95703

SYMPTOMS

  • Extremely poor sensitivity with Xevo TQD
  • It happened after the instrument restart.
  • In TWAVE Diagnostic, the readback of the Collision Ion Guide Pulse Amp becomes a 3-digit value and almost 0 in some places repeatedly.
  • In TWAVE Diagnostic, the readback of HT supply and Collision Guide RF goes up or down repeatedly.
  • Turning off the Operate doesn’t improve.
  • LEDs on the TWAVE generator body are displayed in red.

ENVIRONMENT

  • XevoTQD

CAUSE

Malfunction of TWAVE generator

FIX or WORKAROUND

  1. Improved by unplugging/plugging the LAN cable connected to the bottom

ADDITIONAL INFORMATION


In case of recurrence, replace the 700003510: 48V TWIG ANALYSER UNIT

id95703, XEVOTQD, XEVOTQDIVD

Not able to find a solution? Click here to request help.