Xevo G2-S Argon gas depletes quickly - WKB95489
Article number: 95489
SYMPTOMS
- Xevo G2-S Argon gas depletes quickly
- Tightening all fittings (from the argon cylinder to the instrument, inside the instrument, and where argon gas passes) does not help.
ENVIRONMENT
- Xevo G2-S
CAUSE
Defective argon gas valve
FIX or WORKAROUND
- Replacing the 700009107 GAS ALLY MANIFOLD COMP' INLET 24vDC should be helpful.
ADDITIONAL INFORMATION
id95489, XEVOG2SQTF, XEVOG2STOF