Ebara pump will not pump down after power outage; error "Uu1" - WKB95190
Article number: 95190
SYMPTOMS
- Instrument vented during unplanned power outage
- Error code "Uu1" shows on Ebara front panel, indicating an under voltage status
- Instrument and PC were restarted
- Full instrument communications are established
ENVIRONMENT
- Select Series Cyclic IMS QTof
- Xevo TQ-XS
- Ebara EV-SA30 pump
CAUSE
Latched error code on Ebara pump
FIX or WORKAROUND
- Remove main power from Ebara pump for 10 minutes.
- Reattach main power.
- Pump instrument.
- Verify that instrument pumps properly.
ADDITIONAL INFORMATION
id95190, comms, CYCIMS, EBARA-SA20, EBARA-SA30, XEVOG2XSQT, XEVOTQXS, XEVTQXSIVD