Xevo G2-S QTof does not switch to waste after analysis - WKB94187

Article number: 94187

SYMPTOMS

  • MS diverter valve does not switch to waste after analysis

ENVIRONMENT

  • Xevo G2-S QTof
  • MassLynx

CAUSE

The MS method run time and the time when the valve switches to waste are set exactly the same.

FIX or WORKAROUND

Set the valve switching event to shortly before the end time of the MS method.

ADDITIONAL INFORMATION

id94187, A-SQ, A-SQ, EMD1000, MALDIG2SI, QUATPREMXE, QUMICRO, QUPREMIER, SQD, SQD2, SYNAPT, SYNAPTMS, SYNAPTXS, SYNG2HD, SYNG2HDMAL, SYNG2MS, SYNG2MSMAL, SYNG2SHD, SYNG2SHDML, SYNG2SIHD, SYNG2SIMS, SYNG2SMS, SYNG2SMSML, TQD, TQDIVD, XEVOG2QTOF, XEVOG2SQTF, XEVOG2STOF, XEVOG2TOF, XEVOG2XSQT, XEVOG2XSTF, XEVOQTOF, XEVOTQ, XEVOTQA, XEVOTQD, XEVOTQDIVD, XEVOTQIVD, XEVOTQS, XEVOTQSIVD, XEVOTQSMIC, XEVOTQXS, XEVTQSMIVD, XEVTQXSIVD, ZMD2, ZMD4, ZQ2000, ZQ2000SFC, ZQ4000