Xevo G2-S QTof does not switch to waste after analysis - WKB94187
Article number: 94187
SYMPTOMS
- MS diverter valve does not switch to waste after analysis
ENVIRONMENT
- Xevo G2-S QTof
- MassLynx
CAUSE
The MS method run time and the time when the valve switches to waste are set exactly the same.
FIX or WORKAROUND
Set the valve switching event to shortly before the end time of the MS method.
ADDITIONAL INFORMATION
id94187, A-SQ, A-SQ, EMD1000, MALDIG2SI, QUATPREMXE, QUMICRO, QUPREMIER, SQD, SQD2, SYNAPT, SYNAPTMS, SYNAPTXS, SYNG2HD, SYNG2HDMAL, SYNG2MS, SYNG2MSMAL, SYNG2SHD, SYNG2SHDML, SYNG2SIHD, SYNG2SIMS, SYNG2SMS, SYNG2SMSML, TQD, TQDIVD, XEVOG2QTOF, XEVOG2SQTF, XEVOG2STOF, XEVOG2TOF, XEVOG2XSQT, XEVOG2XSTF, XEVOQTOF, XEVOTQ, XEVOTQA, XEVOTQD, XEVOTQDIVD, XEVOTQIVD, XEVOTQS, XEVOTQSIVD, XEVOTQSMIC, XEVOTQXS, XEVTQSMIVD, XEVTQXSIVD, ZMD2, ZMD4, ZQ2000, ZQ2000SFC, ZQ4000