Sensitivity low for SYNAPT G2 - WKB93040

Article number: 93040

SYMPTOMS

  • Sample signal is significantly lower than expected

ENVIRONMENT

  • SYNAPT G2
  • SYNAPT G2-S
  • SYNAPT G2-Si
  • Xevo G2
  • Xevo G2-S
  • Xevo G2-XS

CAUSE

Electronic communication issue

FIX or WORKAROUND

Reboot electronics and perform detector setup.

ADDITIONAL INFORMATION

Voltages and sliders can hang or become unresponsive if the system has gone a fair amount of time between electronic cycles. Often an electronic reboot clears any latched settings or read backs and is a great starting point when troubleshooting sensitivity or communication issues.