Low sensitivity on TQ-XS with I-Class - WKB92953
Article number: 92953
SYMPTOMS
- Significantly lower sensitivity (to different degrees, depending on the analyte)
- Retention time shift
- Beam for setup solution looks fine
- High pressure difference for SM-FTN Readiness test (up to 2000 psi) or inconsistent results (pressure difference varies in subsequent tests). Pressure difference can also have negative value = ca. -2600 psi)
- Pressure profile for sample syringe leak test is fronting and tailing (see figures below)
Correct pressure profile should look like this:
- Changing position for injection valve from "Inject" to "Load" in interactive display (ACQUITY console) for SM-FTN causes back pressure to increase significantly (should be dropping)
ENVIRONMENT
- Xevo TQ-XS
- ACQUITY I-Class
CAUSE
Partially blocked seat port in SM-FTN
FIX or WORKAROUND
Replace the seat port and the seat port seal in the SM-FTN.
ADDITIONAL INFORMATION