Skip to main content
Waters

Low sensitivity on TQ-XS with I-Class - WKB92953

Article number: 92953

SYMPTOMS

  • Significantly lower sensitivity (to different degrees, depending on the analyte)
  • Retention time shift
  • Beam for setup solution looks fine
  • High pressure difference for SM-FTN Readiness test (up to 2000 psi) or inconsistent results (pressure difference varies in subsequent tests). Pressure difference can also have negative value  = ca. -2600 psi)
  • Pressure profile for sample syringe leak test is fronting and tailing (see figures below)
    EIWTSZ sample syr fronting-tailing.JPG
    Correct pressure profile should look like this:
    EIWTSZ sample syr correct profile.JPG
  • Changing position for injection valve from "Inject" to "Load" in interactive display (ACQUITY console) for SM-FTN causes back pressure to increase significantly (should be dropping)

ENVIRONMENT

  • Xevo TQ-XS
  • ACQUITY I-Class

CAUSE

Partially blocked seat port in SM-FTN

FIX or WORKAROUND

Replace the seat port and the seat port seal in the SM-FTN.

ADDITIONAL INFORMATION

Not able to find a solution? Click here to request help.