Extremely poor sensitivity with Xevo QTof MS - WKB93312
Article number: 93312
SYMPTOMS
- Extremely poor sensitivity noticed on both sample and LockSpray with Xevo QTof MS
- If lower the capillary voltage of the lock spray to about 0.3kV, the sensitivity will increase a little.
- If make the spray direction of the lock spray probe slightly away from the sample cone, the sensitivity will increase a little.
- Replacing everything from the consumable to the FEP red O-ring on the lock spray probe does not help.
ENVIRONMENT
- Xevo QTof MS
CAUSE
Discharge due to dirt on the sample cone.
FIX or WORKAROUND
- Cleaning the sample cone
ADDITIONAL INFORMATION