Extremely poor sensitivity with Xevo QTof MS - WKB93312

Article number: 93312

SYMPTOMS

  • Extremely poor sensitivity noticed on both sample and LockSpray with Xevo QTof MS
  • If lower the capillary voltage of the lock spray to about 0.3kV, the sensitivity will increase a little.
  • If make the spray direction of the lock spray probe slightly away from the sample cone, the sensitivity will increase a little.
  • Replacing everything from the consumable to the FEP red O-ring on the lock spray probe does not help.

ENVIRONMENT

  • Xevo QTof MS

CAUSE

Discharge due to dirt on the sample cone.

FIX or WORKAROUND

  1. Cleaning the sample cone

ADDITIONAL INFORMATION

 

id93312, XEVOG2QTOF, XEVOG2SQTF, XEVOG2STOF, XEVOG2TOF, XEVOG2XSTF, XEVOQTOF