Skip to main content
Waters

ACQUITY BSM stop flow input activated error - WKB92593

Article number: 92593

SYMPTOMS

  • ACQUITY BSM stop flow activated error

ENVIRONMENT

  • ACQUITY UPLC
  • Xevo TQ-S micro
  • MassLynx

CAUSE

The probe inlet tubing has become detached, triggering the leak sensor.

 

FIX or WORKAROUND

  1. Refit the probe inlet tubing.
  2. Dry the leak sensor.

ADDITIONAL INFORMATION

 

id92593, MLYNX, MLYNXV41, SUPMM, UPBINARY, UPBINARY, UPBSM+, UPIBSM, UPIBSM+, UPPBSM, XEVOTQSMIC

Not able to find a solution? Click here to request help.