Frequent loss of communications with mass spectrometer - WKB92094

Article number: 92094

SYMPTOMS

  • Total loss of communications with the MS and UPLC module in the MS console
  • Random communication losses with the MS 
  • Loss of communication with the EPC
  • Possible error in Chronos during the acquisition
  • Errors seen with "Micromass Inlet System Kernel - application error"

ENVIRONMENT

  • MassLynx
  • SYNAPT G2-Si
  • SYNAPT G2-Si MS/HDMS
  • Xevo G2-XS QTof
  • Xevo TQ-XS
  • Any MS

CAUSE

Faulty Ethernet switch box

FIX or WORKAROUND

Replace the network switch box.

ADDITIONAL INFORMATION

 

id92094, A-SQ, A-SQ, comms, LEAPPAL, MALDIG2SI, MLYNX, MLYNXV41, Q3100MSD, QDA, RADIAN, SQD, SQD2, SUPMM, SYNAPT, SYNAPTMS, SYNAPTXS, SYNG2HD, SYNG2HDMAL, SYNG2MS, SYNG2MSMAL, SYNG2SHD, SYNG2SHDML, SYNG2SIHD, SYNG2SIMS, SYNG2SMS, SYNG2SMSML, TQD, TQDIVD, XEVOG2QTOF, XEVOG2SQTF, XEVOG2STOF, XEVOG2TOF, XEVOG2XSQT, XEVOG2XSTF, XEVOG3QTOF, XEVOQTOF, XEVOTQ, XEVOTQA, XEVOTQA, XEVOTQD, XEVOTQDIVD, XEVOTQGC, XEVOTQIVD, XEVOTQS, XEVOTQSCRO, XEVOTQSIVD, XEVOTQSMIC, XEVOTQXS, XEVTQSMIVD, XEVTQXSIVD