BioAccord sensitivity drops - WKB90370

Article number: 90370

SYMPTOMS

  • Drop in sensitivity
  • Running instrument setup does not improve sensitivity

ENVIRONMENT

  • BioAccord System
  • ACQUITY RDa Detector

CAUSE

Dirty aperture disc

FIX or WORKAROUND

Replace the aperture disc (700012106).

ADDITIONAL INFORMATION

id90370, RDA