Article number: 90370

SYMPTOMS

  • Drop in sensitivity
  • Running instrument setup does not improve sensitivity

ENVIRONMENT

  • BioAccord System
  • ACQUITY RDa Detector

CAUSE

Dirty aperture disc

FIX or WORKAROUND

Replace the aperture disc, Part Number (700012106).

O-Ring, FKM, 5.5 ID x 1.6 C/S Part Number 700012105,

ADDITIONAL INFORMATION