When placed into operate the Source temperature does not rise when using Xevo TQ-S micro - WKB90568
Article number: 90568
SYMPTOMS
- Source temperature readback is 24 °C and does not rise when instrument is in Operate mode
- Red LED square found on tune page diagnostics
- Error message when hovering over red LED is "Source heater rise trip, less than 12 °C increase in 3 minutes"
- Resistance across all heater cartridge elements measures 20 ohms
ENVIRONMENT
- Xevo TQ-S micro
CAUSE
Unknown
FIX or WORKAROUND
- Vent the Xevo TQ-S micro.
- Close MassLynx.
- Power-off the Xevo TQ-S micro.
- Wait approximately 1 minute for system to completely power-down.
- Power-on the Xevo TQ-S micro. Allow sufficient time for system to initialize.
- Open MassLynx.
- Pump down the Xevo TQ-S micro.
- Put system in Operate mode.
ADDITIONAL INFORMATION
Note: If system is already vented and off to allow for measuring the resistance across the heater cartridges, follow steps 2 and 5 through 8 after ion block has been reassembled.
id90568, MLYNX, MLYNXV41, SUPMM, XEVOTQSMIC