Skip to main content
Waters

When placed into operate the Source temperature does not rise when using Xevo TQ-S micro - WKB90568

Article number: 90568

SYMPTOMS

  • Source temperature readback is 24 °C and does not rise when instrument is in Operate mode
  • Red LED square found on tune page diagnostics
  • Error message when hovering over red LED is "Source heater rise trip, less than 12 °C increase in 3 minutes"
  • Resistance across all heater cartridge elements measures 20 ohms

ENVIRONMENT

  • Xevo TQ-S micro

CAUSE

Unknown

FIX or WORKAROUND

  1. Vent the Xevo TQ-S micro.
  2. Close MassLynx.
  3. Power-off the Xevo TQ-S micro.
  4. Wait approximately 1 minute for system to completely power-down.
  5. Power-on the Xevo TQ-S micro. Allow sufficient time for system to initialize.
  6. Open MassLynx.
  7. Pump down the Xevo TQ-S micro.
  8. Put system in Operate mode.

ADDITIONAL INFORMATION

Note: If system is already vented and off to allow for measuring the resistance across the heater cartridges, follow steps 2 and 5 through 8 after ion block has been reassembled.

id90568, MLYNX, MLYNXV41, SUPMM, XEVOTQSMIC

Not able to find a solution? Click here to request help.