No beam on TQD - WKB89463
Article number: 89463
SYMPTOMS
- No beam present in positive or negative ion
ENVIRONMENT
- TQ Detector
- Xevo TQ MS
CAUSE
The engineer settings have reverted to default values, resulting in the beam being over-resolved and thus undetectable.
FIX or WORKAROUND
- Load the engineer settings by either uploading the registry settings or reentering the engineer values manually.
- The experimental record from a good data file contains the required engineer setting values.
ADDITIONAL INFORMATION
id89463, TQD, TQDIVD, XEVOTQ, XEVOTQA, XEVOTQIVD