No beam on TQD - WKB89463

Article number: 89463

SYMPTOMS

  • No beam present in positive or negative ion

ENVIRONMENT

  • TQ Detector
  • Xevo TQ MS

CAUSE

The engineer settings have reverted to default values, resulting in the beam being over-resolved and thus undetectable.

FIX or WORKAROUND

  1. Load the engineer settings by either uploading the registry settings or reentering the engineer values manually.
  2. The experimental record from a good data file contains the required engineer setting values.