Skip to main content
Waters

No beam on TQD - WKB89463

Article number: 89463

SYMPTOMS

  • No beam present in positive or negative ion

ENVIRONMENT

  • TQ Detector
  • Xevo TQ MS

CAUSE

The engineer settings have reverted to default values, resulting in the beam being over-resolved and thus undetectable.

FIX or WORKAROUND

  1. Load the engineer settings by either uploading the registry settings or reentering the engineer values manually.
  2. The experimental record from a good data file contains the required engineer setting values.

ADDITIONAL INFORMATION

 

id89463, TQD, TQDIVD, XEVOTQ, XEVOTQA, XEVOTQIVD

Not able to find a solution? Click here to request help.