"Leak Detected" error on MS - WKB89368

Article number: 89368

SYMPTOMS

  • "Leak Detected" error
  • Leak already repaired and sensor dried, but error comes back

ENVIRONMENT

  • Xevo TQ-S micro

CAUSE

Liquid collects between the valve and the front panel, and then drips into the leak tray over time.

FIX or WORKAROUND

  1. Remove the leak sensor.
  2. Stuff a Kimwipe or the like into the tray so it catches drips.
  3. Wait for the tray to dry.

Note: If the system needs to be run before it's completely dry, the check can be disabled in the MS Console under Configure > Intellistart Configuration. Clear the selection for "Leak Detected".

ADDITIONAL INFORMATION

In order to clear the error, it may be necessary to reboot the EPC of the XEVO TQ-S micro. See WKB7776 for instructions on rebooting

 

id89368, XEVOTQSMIC