"Leak Detected" error on MS - WKB89368
Article number: 89368
SYMPTOMS
- "Leak Detected" error
- Leak already repaired and sensor dried, but error comes back
ENVIRONMENT
- Xevo TQ-S micro
CAUSE
Liquid collects between the valve and the front panel, and then drips into the leak tray over time.
FIX or WORKAROUND
- Remove the leak sensor.
- Stuff a Kimwipe or the like into the tray so it catches drips.
- Wait for the tray to dry.
Note: If the system needs to be run before it's completely dry, the check can be disabled in the MS Console under Configure > Intellistart Configuration. Clear the selection for "Leak Detected".
ADDITIONAL INFORMATION
In order to clear the error, it may be necessary to reboot the EPC of the XEVO TQ-S micro. See WKB7776 for instructions on rebooting
id89368, XEVOTQSMIC