Empower QDa: data missing and incorrect mass assignments - WKB8899
Article number: 8899
SYMPTOMS
- "Data Missing" presented in the contour plot when opening a scanning MS channel in the review window
- A typical chromatogram is observed in the channel window
- The spectral data contains incorrect m/z assignments; m/z ratios are out of the operating range of the instrument's capabilities

ENVIRONMENT
- Empower
- QDa
CAUSE
The regional settings on the PC are incorrectly set, which causes the data to be processed incorrectly.
FIX or WORKAROUND
- Ensure that the Windows regional settings are configured for US English (for example, decimal rather than comma):

2. Ensure that the decimal place as symbol is setup correctly in the Windows registry: