Xevo TQS roughing pump vented on its own - WKB8905
Article number: 8905
SYMPTOMS
- roughing pump vented spontaneously
ENVIRONMENT
- Xevo TQ-S
- Xevo TQD
- Xevo TQ
CAUSE
Not enough oil in the roughing pump
FIX or WORKAROUND
- Power off the mass spectrometer.
- Add oil to roughing pump.
- Power on the mass spectrometer.
- Wait three minutes for EPC to fully boot.
- Open MassLynx, tune page, and select pump.
- Let the mass spectrometer pump down for four hours prior to testing (overnight for critical samples).
ADDITIONAL INFORMATION
Part number: 700011138, Synthetic Vacuum Pump Oil