No plates are visible in the MassLynx Plate Generator - WKB86848

Article number: 86848

SYMPTOMS

  • No plates are visible when the Plate Generator is opened from the Inlet Editor (see image in Additional Information)

ENVIRONMENT

  • MassLynx

CAUSE

The Plates folder has been moved, renamed, or deleted from the MassLynx folder on the C drive.

FIX or WORKAROUND

  1. Close MassLynx and its associated windows.
  2. Locate the Plates folder and reinstate it to the correct location in the MassLynx root directory
    C:\MassLynx\Plates
  3. Open MassLynx and ensure that the expected plates are available in the Plate Generator.

ADDITIONAL INFORMATION

plate generator.JPG

Not able to find a solution? Click here to request help.