SYMPTOMS

  • QDa reaches 65% turbo speed and times out 

ENVIRONMENT

  • QDa

CAUSE

Front end gasket leak

FIX or WORKAROUND

  1. Exchange any front end O-rings and gaskets. 
  2. Re-seat the sample cone. 

ADDITIONAL INFORMATION

Not able to find a solution? Click here to request help.