How to measure with ESCi? - WKB85653

Article number: 85653

OBJECTIVE or GOAL

  • Measure with ESCi

ENVIRONMENT

  • SQ Detector 2
  • Xevo TQD
  • Xevo TQ-S
  • Xevo TQ-S micro
  • Xevo TQ-XS
  • Xevo G2-S
  • Xevo G2-XS

PROCEDURE

When using ESCi, perform the following.

  •  Connect the corona pin to the main body as described in your Overview and Maintenance Guide (links to SQ Detector 2 as example)
  • Create an MS Tune file for ESCi.

  Firstly try with the following parameters:

  • Capillary voltage: 1.5 kV
  • Corona current: 10 µA
  • Desolvation temperature: 500℃
  • Desolvation gas: 800L/hr

  •  Change the Source field on the Acquisition tab in MS Method to "API".(Only functions measured in APCI mode)

ADDITIONAL INFORMATION

Not able to find a solution? Click here to request help.