When measuring with APCI, does calibration also need to be performed with APCI? - WKB85648

Article number: 85648

ENVIRONMENT

  • Xevo G2-XS
  • SYNAPT G2-S
  • SYNAPT G2-Si
  • Xevo QTof
  • Xevo G2 QTof
  • Xevo G2-S
  • Xevo G2-XS

ANSWER

Calibration is performed with eg. sodium formate with ESI.

After that, switch to APCI and perform LockSpray Setup and measurement.

ADDITIONAL INFORMATION

 