When measuring with APCI, does calibration also need to be performed with APCI? - WKB85648
Article number: 85648
ENVIRONMENT
- Xevo G2-XS
- SYNAPT G2-S
- SYNAPT G2-Si
- Xevo QTof
- Xevo G2 QTof
- Xevo G2-S
- Xevo G2-XS
ANSWER
Calibration is performed with eg. sodium formate with ESI.
After that, switch to APCI and perform LockSpray Setup and measurement.
ADDITIONAL INFORMATION
id85648, Electrospray