How to calibrate a time-of-flight (TOF) mass spectrometer up to 4000 m/z - WKB8507
Article number: 8507
OBJECTIVE or GOAL
Calibrate TOF MS up to 4000 m/z.
ENVIRONMENT
- Time-of-flight (TOF) mass spectrometer
PROCEDURE
- Infuse NaI or NaCsI.
- Inspect the Scope view in the Tune page to look for large NaI or NaCsI clusters approaching 4000 m/z.
- You may need to increase cone gas flow to increase mass transmission up to 4000 m/z (gas flow of 250 or higher may be necessary).
- You may need to increase cone voltage to increase mass transmission up to 4000 m/z (cone voltage of 100 or higher may be necessary).