Xevo TQ MS error: "Internal error: Reset Instrument from control menu" - WKB84796

Article number: 84796

SYMPTOMS

  • "Internal error: Reset Instrument from control menu"

ENVIRONMENT

  • Xevo TQ MS
  • MassLynx
  • SQ Detector 2
  • Empower

CAUSE

Corrupted tune file selected in Sample Tune Develop Method

FIX or WORKAROUND

Select a different tune file or create a new tune file. 

ADDITIONAL INFORMATION

 