"Calibration expired" error on MS Detector - WKB82522
Article number: 82522
SYMPTOMS
- "Calibration expired" error seen in MS Console
ENVIRONMENT
- 3100 Mass Detector
- SQ Detector
- SQ Detector 2
- TQ Detector
- ACQUITY TQD
- Xevo TQ MS
- Xevo TQD
- Xevo TQ-S
- Xevo TQ-S micro
- Xevo TQ-XS
CAUSE
The calibration file has passed the number of days set on the the IntelliStart configuration for the MS Detector.
FIX or WORKAROUND
- Open the console and open IntelliStart under the MS Detector.
- From the menu, open Configure and choose IntelliStart configuration.
- Highlight the expired calibration file and click Properties.
- For "Test" value, specify the number of days. (Settling the 'Test' value to 365, for example, allows the instrument to work for one year without a calibration error massage.)
ADDITIONAL INFORMATION