Can the source (nitrogen) exhaust and the vacuum pump exhaust from an SQ Detector 2 be vented to the same building exhaust line? - WKB82467

Article number: 82467

ENVIRONMENT

  • SQ Detector 2

ANSWER

Yes, the source (nitrogen) exhaust line and vacuum pump exhaust line can exhausted to the same vent line.

The outlet of the two exhaust lines should be at least 0.5 m apart. If both exhaust lines go into the same trunk line in the wall or ceiling, the end of the vacuum pump exhaust line should extend at least 0.5 m beyond the end of the source (nitrogen) exhaust line.

 