"Error: Scan Error Detected" in Xevo G2-XS (with Hyperterminal) and lost communications with UNIFI - WKB82450
Article number: 82450
SYMPTOMS
- Tof loses communications with UNIFI during long sample sets (intermittent).
- No errors shown in UNIFI.
- If Hyperterminal is collected from LND during run, repeated entries of "Error: Scan Error Detector" appear.
- Electronic restart recovers connection until more long sample sets are run.
ENVIRONMENT
- Xevo G2-XS QTof
- UNIFI 1.9.4 (Server network)
- ACQUITY I-Class UPLC
CAUSE
ADC Data Limit is overloaded by chemical background peaks from LC, causing dropped EPC link.
FIX or WORKAROUND
Perform Acid/Base cleaning on UPLC system.
ADDITIONAL INFORMATION
Recommend following procedure outlined in KB article WKB16167.
Please DO NOT FOLLOW this if you have an Acquity Premier.