"Error: Scan Error Detected" in Xevo G2-XS (with Hyperterminal) and lost communications with UNIFI - WKB82450

Article number: 82450

SYMPTOMS

  • Tof loses communications with UNIFI during long sample sets (intermittent).
  • No errors shown in UNIFI.
  • If Hyperterminal is collected from LND during run, repeated entries of "Error: Scan Error Detector" appear.
  • Electronic restart recovers connection until more long sample sets are run.

ENVIRONMENT

  • Xevo G2-XS QTof
  • UNIFI 1.9.4 (Server network)
  • ACQUITY I-Class UPLC

CAUSE

ADC Data Limit is overloaded by chemical background peaks from LC, causing dropped EPC link.

FIX or WORKAROUND

Perform Acid/Base cleaning on UPLC system.

 

ADDITIONAL INFORMATION

Recommend following procedure outlined in KB article WKB16167.

Please DO NOT FOLLOW this if you have an Acquity Premier. 