Can I connect a QDa and an ELS detector to one Peak NM32LA nitrogen generator? - wkb82382
Article number: 82382
ENVIRONMENT
- ACQUITY QDa Mass Detector
- ACQUITY UPLC ELS Detector
- 2424 Evaporative Light Scattering Detector
- Peak Scientific NM32LA nitrogen generator
ANSWER
Yes
ADDITIONAL INFORMATION
For the QDa, during API operation, typical nitrogen usage is 1300 L/hour (at atmospheric pressure). For the 2424 or ELSD, typical nitrogen usage is 180-240 L/hour. The NM32LA has a maximum output flow rate of approximately 1900 L/hour (32L/min)
See Peak documentation for more information and user manuals.
Use the 6mm Legris tee found in the ELSD startup kit to split the nitrogen between the 2 instruments (700002499 FITTING, TEE, 6MM TUBE TO 6MM TUBE)
For fluidics connection see "How to link an ELSD on an existing LC/MS system - wkb23314".