Outer source heaters are not working on the AutoSpec 1 - WKB80051
Article number: 80051
SYMPTOMS
- Source temperature readback does not reach the set temperature
- No "lights" in the outer source when the heater is requested on; set point is higher than the readback
ENVIRONMENT
- AutoSpec Premier
- AutoSpec
- MassLynx
CAUSE
One or both of the source heaters have failed.
FIX or WORKAROUND
- Verify that the source and analyser ion gauges are on and reading the source vacuum correctly.
- Set the source temperature setting to a high value (for example, 280 °C).
- Confirm that light is emitting from either side of the outer source assembly block.
- Open a Telnet window and inspect the 3481200DC1-S Source Control PCB readbacks:
- Open a Telnet window via the Windows Start/Run menu:
- Type "Telnet epc" into the run open dialog box to start and open a Telnet window.
- Type "rmm" into the Telnet prompt (all lowercase, "r" "m" "m" for rioMainMenu).
- Specify "3" to access subsystems, specify "4" to access the 3481200DC1-S Source Control PCB, and then specify "3" to display a snapshot of the source control PCB readbacks.
- Set the source to a temperature higher than the readback and type "3" into the Telnet window to list the readbacks when the heater should be on.
- A working source heater will show the following:
- 1503. SourceHeaterVoltageRb 7.254902 (actual value may differ depending on the temperature difference between the set point and the readback)
- An open-circuit source heater will show the following:
- 1503. SourceHeaterVoltageRb 24.xxx
- A working source heater will show the following:
- Set the source to 0 and type "3" into the Telnet window to list the readbacks when the heater is off.
- A working source heater will show the following:
- 1503. SourceHeaterVoltageRb 0.xxx
- A working source heater will show the following:
- The readbacks above show the 3481200DC1-S Source Control PCB to be electrically functioning, suggesting an open-circuit source heater.
- Type "999" into the Telnet box repeatedly until the -> prompt is seen.
- Close the Telnet window.
- Vent the source region per standard system operating procedures and remove the outer source assembly.
- Using a digital voltmeter set to resistance measurement (0 to 10 Ohms if the scale is manually set), test the short circuit resistance of the meter leads:
- Touch the test leads together and note the resistance: expect 0 to 0.5 Ohms.
- If the reading is higher, try new test leads or an alternative digital voltmeter.
- Referring to the Outer Source technical reference, circuit measure the resistance through the outer source heater circuit.
- Measure resistance between pins 7 and 8 on the left-hand side contact rail.
- Expect a resistance of 0.5 to 1.5 Ohms.
- If a high or open circuit resistance is measured, one or both of the heaters has failed and will need replacement.
- It is good practice to change both heaters at the same time. Order two of M702867AD1 - Source Heater.
- Ensure that no shorts are present on the heater connection at the insulated/terminal connection end.
- Finalize the build with a repeat resistance measurement as above, and measure the resistance between the ion block and the Focus 1 lens plate.
- Expect an open circuit, infinite resistance.
- If a low/short resistance is measured, check for shorts on the filament shields between the ion block and the Focus 1 lens.
ADDITIONAL INFORMATION
The Outer Source will show a raised temperature due to heating from the GC Interface and when in Operate, from the Source filament. Temperatures may rise up to 180 °C.
See article WKB118407 on procedure on replacing the outer source heater. It is highly recommended that both heaters are replaced at the same time.
See related article wkb281938.
