Xevo TQ MS has no beam on the tune page - WKB79579
Article number: 79579
SYMPTOMS
- Intensity of the Tune page is fluctuating so rapidly that it cannot be read
- Infusion parameters are set up correctly
- Flow is coming from the probe into the instrument
ENVIRONMENT
- Xevo TQ MS
- Xevo TQD
- Xevo TQ-S
- Xevo TQ-S micro
- Xevo TQ-XS
CAUSE
The Scan Time parameter is set too low, causing the instrument to fluctuate between channels too rapidly.
FIX or WORKAROUND
- On the MS Tune Page, select Setup.
- Open Scope Parameters.
- The default values should be set as follows:
- Scan Time (s) = 0.1
- Inter-scan Delay (s) = 0.02
- Press OK.
ADDITIONAL INFORMATION