Flow profile on neonatal system is unstable - WKB74983
Article number: 74983
SYMPTOMS
- The flow profile starts off normal but becomes unstable
ENVIRONMENT
- TQ Detector
- 2777 Sample Manager
- 1525 Micro Binary HPLC Pump
CAUSE
The rotor seal is wearing out on the 2777 and causing a partial blockage in the ESI probe of the TQD. This has the effect of making the beam unstable when the flow profile drops to a low flow rate.
FIX or WORKAROUND
- Replace the rotor seal on the 2777.
- Rebuild the ESI probe.
ADDITIONAL INFORMATION
Rotor seal (700002210)
ESI probe rebuild kit (700005795)
