Flow profile on neonatal system is unstable - WKB74983

Article number: 74983

SYMPTOMS

  • The flow profile starts off normal but becomes unstable

ENVIRONMENT

  • TQ Detector
  • 2777 Sample Manager
  • 1525 Micro Binary HPLC Pump

CAUSE

The rotor seal is wearing out on the 2777 and causing a partial blockage in the ESI probe of the TQD. This has the effect of making the beam unstable when the flow profile drops to a low flow rate.

FIX or WORKAROUND

  1. Replace the rotor seal on the 2777.
  2. Rebuild the ESI probe.

ADDITIONAL INFORMATION

Rotor seal (700002210)

ESI probe rebuild kit (700005795)

id74983, 2777, 2777C, 2777IVD, Electrospray, TQD, TQDIVD