Mass spectrometer is present in DHCP server but shows no communications in MassLynx - WKB7423
Article number: 7423
SYMPTOMS
- No communication with MS
- "Instrument not present" shown in MassLynx
- DHCP server does detect the instrument and shows it has an IP address assigned
- Opening a command prompt and typing "Ping EPC" results in "no reply from" message
- Opening a command prompt and pinging the IP address listed in the DHCP server brings no reply from the EPC
ENVIRONMENT
- 3100 Mass Detector
- ACQUITY QDa Mass Detector
- ACQUITY UPLC
- ACQUITY UPLC H-Class
- ACQUITY UPLC I-class
- MassLynx
- SQ Detector 2
- SYNAPT G2
- SYNAPT G2-S
- SYNAPT G2-Si
- SYNAPT XS
- TQ Detector
- Xevo QTof
- Xevo G2
- Xevo G2-S
- Xevo G2-XS
- Xevo TQ MS
- Xevo TQD
- Xevo TQ-S
- Xevo TQ-S micro
- Xevo TQ-S cronos
- Xevo TQ-XS
CAUSE
Mismatch of IP address assigned by the Waters DHCP server and the hosts file (C:\Windows\System32\drivers\etc\hosts), for example:
- Two IP addresses listed for the EPC, and the top one is not that shown in the DHCP server
- Two IP addresses listed for the EPC, neither of which is the same as the one shown in the DHCP server
- Single IP address listed for the EPC, which is not the same as the one shown in the DHCP server
FIX or WORKAROUND
- Close MassLynx.
- Open the hosts file (C:\Windows\System32\drivers\etc\hosts) using Notepad.
- Edit the hosts file so that the IP address assigned to the EPC matches the one listed in the DHCP server (example shown below).
-
- Open a command prompt and type Ping EPC to confirm that the EPC is now associated with the correct IP address.
- If Ping EPC works, reopen MassLynx and see if the instrument is now present.
ADDITIONAL INFORMATION