Xevo TQ fails source pressure test with exhaust trip warning - WKB7378
Article number: 7378
SYMPTOMS
- MS fails source pressure test with an exhaust trip warning
ENVIRONMENT
- Xevo TQ
CAUSE
Restricted API gas feed-through line in the source enclosure
FIX or WORKAROUND
- Open the source enclosure door.
- Locate the tubing that feeds the API gas into the source enclosure (feeds through small green gas O-ring/ seal).
- Blow compressed air into the feed line to alleviate the restriction.
- Close the source enclosure door and wait for the source pressure test to successfully complete.
ADDITIONAL INFORMATION
