Poor reproducibility for negative transitions with Xevo TQ-S - WKB73040
Article number: 73040
SYMPTOMS
- Poor reproducibility for negative transitions
- Positive transitions all acceptable
ENVIRONMENT
- Xevo TQ-S
- Quattro Premier XE
CAUSE
Large number of MRMs set up in positive and negative modes in the same time span of MS method
FIX or WORKAROUND
- Separate MRM groups to minimize MRMs overlapping with negative retention time window.
- Validate new MS method to ensure that positive and negative data are reproducible and acceptable.