Poor sensitivity after power outage on TQD - WKB73181
Article number: 73181
SYMPTOMS
- No peaks after power outage
- Increase of collision energy to 20V brings peaks back, but sensitivity is still diminished
- All tune readbacks appear correct
- There is no difference between MS1, MS2, and MSMS, in terms of sensitivity
- Collision guide output voltage readbacks on the T-Wave diagnostics page appears to be 20V when collision energy is set to 0V
ENVIRONMENT
- ACQUITY TQD
- MassLynx
CAUSE
A 20V offset voltage is being applied to the collision ion guide output voltage. Setting Collision Energy to 0V should give a collision ion guide output voltage readback of 0V.
FIX or WORKAROUND
- Cycle power to the instrument.
- If voltage offset is still present after cycling power, replace the ACQUITY TQD T-Wave Generator (700003510).
ADDITIONAL INFORMATION
A collision energy offset can be noted on the "T-Wave Diagnostic" tab in the "Collision Ion Guide Pulse Amp" section. The reading should be opposite polarity of the set collision energy. Example: If collision energy is set to 20V in daughter mode, the average "Collision Ion Guide Pulse Amp" reading should be around -20V.
The collision energy offset noted above in the "Cause" section may be something higher or lower than 20V (for example, 15V).
