High background noise on Xevo TQ-S micro; fails sensitivity specification - WKB73035
Article number: 73035
SYMPTOMS
- Fails sensitivity specification
- High background noise
ENVIRONMENT
- SQ Detector
- SQ Detector 2
- TQ Detector
- Xevo TQD
- Xevo TQ-S cronos
- Xevo TQ-S micro
- Xevo TQ MS
- Xevo TQ-S
- Xevo TQ-XS
- SNYAPT G2
- SNYAPT G2-S
- SYNAPT G2-Si
- Xevo G2
- Xevo G2-S
- Xevo G2-XS
- Xevo QToF
CAUSE
Poor solvent quality
FIX or WORKAROUND
- Make up the solvents using our solvent kits (186008715).
- Be sure to use formic acid from a glass ampule and not a plastic bottle.
ADDITIONAL INFORMATION
id73035, A-SQ, A-SQ, MALDIG2SI, SQD, SQD2, SYNAPT, SYNAPTMS, SYNG2HD, SYNG2HDMAL, SYNG2MS, SYNG2MSMAL, SYNG2SHD, SYNG2SHDML, SYNG2SIHD, SYNG2SIMS, SYNG2SMS, SYNG2SMSML, TQD, TQDIVD, XEVOG2QTOF, XEVOG2SQTF, XEVOG2STOF, XEVOG2TOF, XEVOG2XSQT, XEVOG2XSTF, XEVOQTOF, XEVOTQ, XEVOTQA, XEVOTQD, XEVOTQDIVD, XEVOTQIVD, XEVOTQS, XEVOTQSCRO, XEVOTQSIVD, XEVOTQSMIC, XEVOTQXS, XEVTQSMIVD, XEVTQXSIVD