Skip to main content
Waters

No signal with TQD - WKB71621

Article number: 71621

SYMPTOMS

  • No signal
  • Spray present

ENVIRONMENT

  • TQ Detector

CAUSE

Parameters on the tune page are not adjusted correctly.

FIX or WORKAROUND

  1. Inspect all parameters on each tab of the tune page.
  2. In this case, ion energy 1 and ion energy 2 were adjusted at -11.2eV.
  3. Readjust each ion energy at approximately 0.5eV.

ADDITIONAL INFORMATION

Not able to find a solution? Click here to request help.