Gradually decreased signal seen when successive injections of the same standard are performed on an ACQUITY UPLC system connected to Xevo TQ MS - WKB71602
Article number: 71602
SYMPTOMS
- Gradually decreased signal (as low as <50% of the initial level) is seen when successive injections of the same standard are performed on an ACQUITY UPLC system connected to Xevo TQ MS Detector
- The signal returns to the initial level after priming the SM
ENVIRONMENT
- ACQUITY UPLC SM
- Xevo TQ MS
CAUSE
Non-pre-slit vials were used, resulting in gradually decreased signal in successive injections of the same vial.
FIX or WORKAROUND
Use pre-slit vials or other pre-slit sample containers.
ADDITIONAL INFORMATION
id71602, UPBINARY, XEVOTQ, XEVOTQA, XEVOTQIVD