Pump HW overpressure error on LCMS systems - WKB69828
Article number: 69828
SYMPTOMS
- Pump pressure is normal when equilibrating system
- Pump HW overpressure error occurs within a short time of starting a run
ENVIRONMENT
- 3100 Mass Detector
- SQ Detector
- SQ Detector 2
- TQ Detector
- Xevo TQ MS
- Xevo TQD
- Xevo TQ-GC
- Xevo TQ-S
- Xevo TQ-S micro
- Xevo TQ-XS
CAUSE
The tubing from the diverter valve to the probe is blocked or kinked. When the diverter valve switches from "waste" to "LC" during a run, the overpressure error occurs.
FIX or WORKAROUND
Replace the tubing with part 700003799 (Black PEEK 1/16" x 0.004 x 60").
ADDITIONAL INFORMATION
id69828, A-SQ, A-SQ, Q3100MSD, SQD, SQD2, TQD, TQDIVD