ESI or UniSpray source pressure test fails - WKB6822
Article number: 6822
SYMPTOMS
- The pressure test fails when a ESI or UniSpray source is fitted to the instrument
- Gas leaking from the source may be audible
ENVIRONMENT
- LockSpray Source
- UniSpray Source
- Xevo TQ Absolute
- Xevo TQ-XS
- Xevo TQ-S
- Xevo TQD
- Xevo TQ-S micro
- Xevo G2
- Xevo G2-S
- Xevo G2-XS
- Synapt G2
- Synapt G2-S
- Synapt G2-Si
- Vion IMS QTOF
CAUSE
Incorrectly seated O-rings in the source door
FIX or WORKAROUND
- Reseat or replace the source enclosure seal (part no. 700004320) and the two small gas seals (part no. 700004322) in the source door.
ADDITIONAL INFORMATION
id6822, Electrospray, MALDIG2SI, SYNG2HD, SYNG2HDMAL, SYNG2MS, SYNG2MSMAL, SYNG2SHD, SYNG2SHDML, SYNG2SIHD, SYNG2SIMS, SYNG2SMS, SYNG2SMSML, VION, XEVOG2QTOF, XEVOG2SQTF, XEVOG2STOF, XEVOG2TOF, XEVOG2XSQT, XEVOG2XSTF, XEVOQTOF, XEVOTQ, XEVOTQA, XEVOTQD, XEVOTQDIVD, XEVOTQIVD, XEVOTQS, XEVOTQSIVD, XEVOTQSMIC, XEVOTQXS, XEVTQSMIVD, XEVTQXSIVD