Skip to main content
Waters

ESI or UniSpray source pressure test fails - WKB6822

Article number: 6822

SYMPTOMS

  • The pressure test fails when a ESI or UniSpray source is fitted to the instrument
    clipboard_eb55d09fa664401248b13d4a8972846a6.png
  • Gas leaking from the source may be audible

ENVIRONMENT

  • LockSpray Source
  • UniSpray Source
  • Xevo TQ Absolute 
  • Xevo TQ-XS
  • Xevo TQ-S
  • Xevo TQD
  • Xevo TQ-S micro
  • Xevo G2
  • Xevo G2-S
  • Xevo G2-XS
  • Synapt G2
  • Synapt G2-S
  • Synapt G2-Si
  • Vion IMS QTOF

CAUSE

Incorrectly seated O-rings in the source door

FIX or WORKAROUND

  1. Reseat or replace the source enclosure seal (part no. 700004320) and the two small gas seals (part no. 700004322) in the source door.

ADDITIONAL INFORMATION

Not able to find a solution? Click here to request help.