UniSpray source pressure test fails - WKB6822

Article number: 6822

SYMPTOMS

  • The pressure test fails when a ESI or UniSpray source is fitted to the instrument
  • Gas leaking from the source may be audible.

ENVIRONMENT

  • LockSpray Source
  • UniSpray Source
  • Xevo TQ-XS
  • Xevo TQ-s
  • Xevo G2
  • Xevo G2-S
  • Xevo G2-XS
  • Synapt G2
  • Synapt G2-S
  • Synapt G2-Si
  • Vion IMS QTOF

CAUSE

Incorrectly seated O-rings in the source door

FIX or WORKAROUND

  1. Reseat or replace the source enclosure seal and the two small gas seals in the source door.

ADDITIONAL INFORMATION

 