SQ Detector 2 fails source pressure test - WKB66909
Article number: 66909
SYMPTOMS
- Source pressure test failure
- System was vented to change vacuum pump oil
- System pumped down, turbo speed 100%
- Opening and closing source housing to rerun test does not resolve
- Recently had issue with exhaust tubing becoming disconnected
- Reconnecting tubing resolves original issue
- Exhaust tubing found to be connected
ENVIRONMENT
- SQ Detector 2
CAUSE
Likely issue is leak in exhaust tubing or failed valve on trap bottle.
FIX or WORKAROUND
On Tune page, go to Gas > Override Source Pressure Test.
ADDITIONAL INFORMATION
Note: Customer was looking for workaround to run samples right away. Replacement of valve, tubing, or fittings may be required.
id66909, A-SQ, SQD2