Poor Tof mass accuracy - WKB65254
Article number: 65254
SYMPTOMS
- Poor mass accuracy on Tof
ENVIRONMENT
- Waters Tof system
- Vion IMS QTof
CAUSE
System calibration and lockspray setup have not been completed.
FIX or WORKAROUND
- Check mass accuracy using leucine enkephalin.
- Optimize peak resolution.
- Adjust Veff to the accurate mass for the leucine enkephalin peak and save new settings.
- Complete the calibration and lockspray setup.