Sensitivity loss on MassLynx V4.1 using a Xevo TQ-S Micro - WKB6458
Article number: 6458
SYMPTOMS
- Large sensitivity drop overnight
ENVIRONMENT
- MassLynx
- Xevo TQ-S Micro
CAUSE
Incorrect tune page loaded
FIX or WORKAROUND
- Ensure that the correct tune page is open.
- Ensure that the correct tune page is specified in the sample list.
ADDITIONAL INFORMATION
Comparing the experimental records of good and bad injections is the best way to determine if acquisition parameters differ.