What are the nitrogen gas requirements for a Xevo G2 QTof? - WKB64481

Article number: 64481

ENVIRONMENT

  • Xevo G2 QTof
  • Xevo G2 Tof

ANSWER

The Xevo G2 QTof requires a supply of dry, oil-free nitrogen with a purity of at least 95%. The nitrogen must be regulated at 7 bar (100 psi) outlet pressure. During API operation, typical nitrogen usage varies from 600 to 1200 L/h.

ADDITIONAL INFORMATION

If an APGC source is present, the nitrogen purity must be >99.999%.