Skip to main content
Waters

How to remove the pumping block insert assembly on the SQD2, Xevo TQ-S micro or the QDa - WKB6442

Article number: 6442

OBJECTIVE or GOAL

Remove the Xevo TQ-S micro pumping block insert assembly. (When you try to clean the Stepwave on a Xevo TQ-S Micro, the pumping block insert assembly may become stuck.)

ENVIRONMENT

  • Xevo TQ-S Micro
  • ACQUITY QDa Mass Detector
  • SQ Detector 2

PROCEDURE

  1. Remove the cone.
  2. Remove the four holding screws for the pumping block insert assembly.
  3. Start to wiggle the assembly free.
  4. Apply methanol to the edges of the screw holes and around the edges of the pumping block.
  5. Continue to wiggle the assembly free.

ADDITIONAL INFORMATION

Be sure to leave the PEEK block and preferably the ion block (minus the cone) on so as to have something to grip when removing the assembly.

See Section 4.18 (Rev. C) of Xevo TQ-S micro Overview and Maintenance Guide, 715004599

See Section 3.10 (Rev. F) of ACQUITY QDa Detector Overview and Maintenance Guide, 715003956

Not able to find a solution? Click here to request help.