No capillary voltage on MS tune page when MS is in operate - WKB64398
Article number: 64398
SYMPTOMS
- Capillary voltage stays at 0 KV on ES+
- All other parameters are okay
ENVIRONMENT
- SQD2
- Xevo TQD
- 3100
- MassLynx
CAUSE
Electronical failure due to a power failure in the lab
FIX or WORKAROUND
- Switch on ES- for 10 seconds.
- Switch on ES+.
ADDITIONAL INFORMATION