Poor Tof MRM sensitivity - WKB6335
Article number: 6335
SYMPTOMS
- Tof MRM experiment is less sensitive than an MSe experiment with the same compound
ENVIRONMENT
- Xevo G2-XS
- SYNAPT G2-S
- SYNAPT G2-Si
- Tof MRM
CAUSE
The Tof MRM experiment is not set up properly.
FIX or WORKAROUND
In the Tof MRM experiment:
- Do not select RADAR.
- Be sure to select Single Isotope or Isotope Cluster for the MRM window.
In the MRM Editor window:
- Select a single collision energy (CE) if the best collision energy is known.
- Set the Target Enhancement m/z for the main daughter ion of interest.
ADDITIONAL INFORMATION
For information on how to set up a Tof MRM experiment, see the article How to setup a Tof MRM experiment.