Incorrect collision gas flow resulting in lower sensitivity on ACQUITY TQD - WKB63254

Article number: 63254

SYMPTOMS

  • Low sensitivity in MSMS mode

ENVIRONMENT

  • ACQUITY TQD
  • TQ Detector

CAUSE

Incorrect collision gas flow

FIX or WORKAROUND

  1. Correct the collision gas flow so that collision cell pressure is between 3.5e-3 and 4.0e-3.
  2. Verify that sensitivity of a known method is as expected.

ADDITIONAL INFORMATION

id63254, TQD, TQDIVD