Sensitivity dropping over time - WKB62583
Article number: 62583
SYMPTOMS
Sensitivity dropping over time.
ENVIRONMENT
- Xevo G2-XS
- MassLynx
CAUSE
Automatic detector gain control not enabled.
FIX or WORKAROUND
Enable automatic detector gain control.
ADDITIONAL INFORMATION
A polarity switch did not restore the sensitivity.
The Automatic Detector Check software monitors and adjusts the detector voltage to maintain a steady detector gain when running ESI and nano-ESI experiments, and flags the need to run the detector setup.