In the MassLynx calibration file for a SYNAPT G2, the ESI NaICS Pos reference file is missing information - WKB61999
Article number: 61999
SYMPTOMS
- Opened in Notepad, the ESI_NaICs_Pos.ref file doesn't show how much sodium to add:
- There is note that use with automatic calibration is not recommended, but no reason is given
ENVIRONMENT
- SYNAPT G2
- MassLynx
- ESI_NaICs_Pos.ref file
CAUSE
Unknown
FIX or WORKAROUND
Use the recipe as given for makeup of the API Calibration solution (700001593), which contains 2 µg/uL of sodium iodide and 50 ng/uL of cesium iodide in 50:50 propan-2-ol/water.
ADDITIONAL INFORMATION
Automatic calibrations are not recommended because the software can match the wrong peak. It is better to verify the peak matching process.