In the MassLynx calibration file for a SYNAPT G2, the ESI NaICS Pos reference file is missing information - WKB61999
Article number: 61999
SYMPTOMS
- Opened in Notepad, the ESI_NaICs_Pos.ref file doesn't show how much sodium to add:
- There is note that use with automatic calibration is not recommended, but no reason is given
ENVIRONMENT
- SYNAPT G2
- MassLynx
- ESI_NaICs_Pos.ref file
CAUSE
Unknown
FIX or WORKAROUND
Use the recipe as given for makeup of the API Calibration solution (700001593), which contains 2 µg/uL of sodium iodide and 50 ng/uL of cesium iodide in 50:50 propan-2-ol/water.
ADDITIONAL INFORMATION
Automatic calibrations are not recommended because the software can match the wrong peak. It is better to verify the peak matching process.
id61999, Electrospray, MLYNX, MLYNXV41, SUPMM, SYNG2HD, SYNG2HDMAL, SYNG2MS, SYNG2MSMAL