Skip to main content
Waters

In the MassLynx calibration file for a SYNAPT G2, the ESI NaICS Pos reference file is missing information - WKB61999

Article number: 61999

SYMPTOMS

  • Opened in Notepad, the ESI_NaICs_Pos.ref file doesn't show how much sodium to add:ref.JPG
  • There is note that use with automatic calibration is not recommended, but no reason is given

ENVIRONMENT

  • SYNAPT G2
  • MassLynx
  • ESI_NaICs_Pos.ref file

CAUSE

Unknown

FIX or WORKAROUND

Use the recipe as given for makeup of the API Calibration solution (700001593), which contains 2 µg/uL of sodium iodide and 50 ng/uL of cesium iodide in 50:50 propan-2-ol/water.

ADDITIONAL INFORMATION

Automatic calibrations are not recommended because the software can match the wrong peak. It is better to verify the peak matching process. 

Not able to find a solution? Click here to request help.