With QDa, cannot set scan range to 200-1250 Da - WKB61872
Article number: 61872
SYMPTOMS
- When the scanning range of the Empower method is set to 200-1250 Da, the instrument status is always Running and cannot be controlled
ENVIRONMENT
- ACQUITY QDa Mass Detector
- Empower
CAUSE
The method setting exceeds the upper limit of the instrument scan number.
FIX or WORKAROUND
- When the Empower method scan range is set to 200-1250 Da, the accuracy rate should be less than 10 (recommendation: 8).
- Recommendation: Set the Empower method scan range to 200-1200 Da.
ADDITIONAL INFORMATION
