Xevo G2-XS fails source pressure test - WKB58720

Article number: 58720

SYMPTOMS

  • Software reports that the source pressure test is failing
  • No vacuum readbacks are showing
  • Electronics reboot does not resolve issue

ENVIRONMENT

  • MassLynx
  • Xevo G2-XS

CAUSE

Communication loss with the system

FIX or WORKAROUND

Perform an electronics reboot, but wait longer (several minutes) before starting MassLynx.

ADDITIONAL INFORMATION